Quantity One User Guide
5-10
Save After Scan, the file will not be saved until you select Save or Save As
from the File menu.
5.6 Calibration
If you have installed a calibration overlay, you can automatically calibrate
your transmissive and reflective scans. (Calibration overlays for the GS-700
can be ordered from Bio-Rad.)
To set the automatic calibration settings, click on the More Options button in
the GS-700 acquisition window. This will open the Densitometer Options
dialog box.
Fig. 5-9. More Options in the GS-700 acquisition window.
To enable automatic calibration, click on the Calibration On checkbox.
Note: Before you can calibrate, you must enter the correct step tablet values for
your transmissive calibration strip into the Step Tablet Values form, as
Vista de pagina 91