GS-710 Imaging Densitometer
6-11
step tablet to which those density values belong. Those exact density values
must be entered into the computer for the software to associate a correct
density value with each step on the step tablet.
Note: Scanning in transmissive mode with incorrect step tablet values entered
into the computer can cause significant errors in the reported densities of
your scans.
To enter the density values for your step tablet, click on the Edit Step Tablet
button in the Densitometer Options box. The Step Tablet Values dialog box for
the GS-710 will open.
Fig. 6-10. Step Tablet dialog box.
The dialog box will indicate whether the values are for the transmissive or
reflective step tablet, depending on whether you are in transmissive or
reflective mode.
If you are in reflective mode, you do not need to change the default step tablet
values.